3 Installation
Proline Promag 51
38
Hauser
3.4
Post-installation check
Perform the following checks after installing the measuring device in the pipe:
Device condition and specifications
Notes
Is the device damaged (visual inspection)?
−
Does the device correspond to specifications at the measuring point, including
process temperature and pressure, ambient temperature, minimum fluid conduc-
tivity, measuring range, etc.?
Installation
Notes
Does the arrow on the sensor nameplate match the direction of flow through
the pipe?
−
Is the plane of the measuring electrode axis correct?
Horizontal?
Is the position of the Empty Pipe Detection (EPD) electrode correct?
Were all threaded fasteners tightened to the specified torques when the sensor
was installed?
Were the correct seals installed (type, material, installation)?
Promag W
→
Promag P
→
Are the measuring point number and labeling correct (visual inspection)?
−
Process environment / process conditions
Notes
Are the inlet and outlet runs to respected?
Inlet run
≥
5 x DN
Outlet run
≥
2 x DN
Is the measuring device protected against moisture and direct sunlight?
−
Is the sensor adequately protected against vibration (attachment, support)?
Acceleration up to 2 g by analogy
with IEC 600 68-2-6
(see Page 109)
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...