Troubleshooting
Promag 50
92
Hauser
9.6.3
Replacing the exchangeable electrode
The Promag W sensor (DN 350 to 2000; 14" to 78") is available with exchangeable measuring
electrodes as an option. This design permits the measuring electrodes to be replaced or cleaned
under process conditions.
a0004447
Fig. 56:
Apparatus for replacing exchangeable measuring electrodes
View A = DN 1200 to 2000 (48" to 78")
View B = DN 350 to 1050 (14" to 42")
1
Allen screw
2
Handle
3
Electrode cable
4
Knurled nut (locknut)
5
Measuring electrode
6
Stop cock (ball valve)
7
Retaining cylinder
8
Locking pin (for handle)
9
Ball-valve housing
10
Seal (retaining cylinder)
11
Coil spring
A
3
2
1
4
8
7
11
1
6
10
9
5
B