Troubleshooting
Promag 50
82
Hauser
9.3
Process error messages
!
Note!
Also observe the information on
61.
No. # 5xx
Application error
501
S: SW.-UPDATE ACT.
!: # 501
New amplifier or communication (I/O module)
software version is loaded.
Currently no other functions are possible.
Wait until the procedure is finished.
The device will restart automatically.
502
S: UP-/DOWNLOAD ACT
!: # 502
Uploading or downloading the device data via operating
program.
Currently no other functions are possible.
Wait until the procedure is finished.
No. # 6xx
Simulation mode active
601
S: POS. ZERO-RETURN
!: # 601
Positive zero return active
"
Caution!
This message has the highest display priority!
Switch off positive zero return
611
to
614
S: SIM. CURR. OUT. n
!: # 611 to 614
Simulation current output active
621
to
624
S: SIM. FREQ. OUT. n
!: # 621 to 624
Simulation frequency output active
Switch off simulation
631
to
634
S: SIM. PULSE n
!: # 631 to 634
Simulation pulse output active
Switch off simulation
641
to
644
S: SIM. STAT. OUT n
!: # 641 to 644
Simulation status output active
Switch off simulation
671
to
674
S: SIM. STATUS IN n
!: # 671 to 674
Simulation status input active
Switch off simulation
691
S: SIM. FAILSAFE
!: # 691
Simulation of response to error (outputs) active
Switch off simulation
692
S: SIM. MEASURAND
!: # 692
Simulation of a measured variable active (e.g. mass
flow).
Switch off simulation
698
S: DEV. TEST ACT.
!: # 698
The measuring device is being checked on site via the
test and simulation device.
–
No.
Error message / Type
Cause
Remedy (spare part
ä
86)
No.
Error message / Type
Cause
Remedy (spare part
P = Process error
$
= Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
401
EMPTY PIPE
$
: # 401
Measuring tube partially filled or empty
1.
Check the process conditions of the plant
2.
Fill the measuring tube
461
ADJ. NOT OK
!: # 461
EPD calibration not possible because the fluid's
conductivity is either too low or too high.
The EPD function cannot be used with fluids of this nature.
463
FULL = EMPTY
$
: # 463
The EPD calibration values for empty pipe and full pipe
are identical, therefore incorrect.
Repeat calibration, making sure procedure is correct