Description of device parameters
Proline Promag 300 Modbus RS485
208
Hauser
Description
Use this function to switch simulation of the switch output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection
• Off
• On
Factory setting
Off
Additional information
Description
The desired simulation value is defined in the
Switch status 1 to n
parameter.
Selection
• Off
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Switch simulation is active.
Switch status 1 to n
Navigation
Expert → Diagnostics → Simulation → Switch status 1 to n (0463–1 to n)
Description
Use this function to select a switch value for the simulation. In this way, users can verify
the correct adjustment of the switch output and the correct function of downstream
switching units.
Selection
• Open
• Closed
Additional information
Selection
• Open
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• Closed
Switch simulation is active.
Relay out. 1 to n sim
Navigation
Expert → Diagnostics → Simulation → Relay out. 1 to n sim (0802–1 to n)
Description
Use this function to switch simulation of the relay output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection
• Off
• On
Factory setting
Off