Proline Cubemass C 500
96
Hauser
Supported operating
tools
Operating unit
Interface
Additional information
FieldCare SFE500
Notebook, PC or tablet
with Microsoft Windows
system
• CDI-RJ45 service
interface
• WLAN interface
• Fieldbus protocol
Device Xpert
Field Xpert SFX
100/350/370
HART and
FOUNDATION Fieldbus
fieldbus protocol
Operating Instructions
BA01202S
Device description files:
Use update function of
handheld terminal
Other operating tools based on FDT technology with a device driver such as DTM/iDTM or
DD/EDD can be used for device operation. These operating tools are available from the
individual manufacturers. Integration into the following operating tools, among others, is
supported:
• FactoryTalk AssetCentre (FTAC) by Rockwell Automation →
www.rockwellautomation.com
• Process Device Manager (PDM) by Siemens →
www.siemens.com
• Asset Management Solutions (AMS) by Emerson →
www.emersonprocess.com
• FieldCommunicator 375/475 by Emerson →
www.emersonprocess.com
• Field Device Manager (FDM) by Honeywell →
www.honeywellprocess.com
• FieldMate by Yokogawa →
www.yokogawa.com
• PACTWare →
www.pactware.com
The associated device description files are available at:
www.endress.com
→ Downloads
Web server
Thanks to the integrated Web server, the device can be operated and configured via a Web browser
and via a service interface (CDI-RJ45) or via a WLAN interface. The structure of the operating menu
is the same as for the local display. In addition to the measured values, status information on the
device is also displayed and allows the user to monitor the status of the device. Furthermore the
device data can be managed and the network parameters can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the WLAN
connection: order code for "Display; operation", option
G
"4-line, illuminated; touch c WLAN".
The device acts as an Access Point and enables communication by computer or a mobile handheld
terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the measuring
device:
– Upload the configuration from the measuring device (XML format, configuration backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file or PDF file, document the measuring point configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat Verification"
application package)
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
– Visualize up to 1000 saved measured values (only available with the
Extended HistoROM
application package → 103)
Web server special documentation → 108
HistoROM data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making operation and
servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored as a
backup in the device memory. This memory can be overwritten with an updated data record, for
example after commissioning.