Proline Cubemass C 100 HART
Commissioning
Hauser
61
Parameter
Description
Selection / User entry
Factory setting
Mass flow unit
Select mass flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit choose list
Country-specific:
• kg/h
• lb/min
Volume flow unit
Select volume flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit choose list
Country-specific:
• l/h
• gal/min (us)
Minimum frequency value
Enter minimum frequency.
0.0 to 10
000.0 Hz
0.0 Hz
Maximum frequency value
Enter maximum frequency.
0.0 to 10
000.0 Hz
10
000.0 Hz
Measuring value at minimum
frequency
Enter measured value for minmum
frequency.
Signed floating-point number
0
Measuring value at maximum
frequency
Enter measured value for maximum
frequency.
Signed floating-point number
0
Failure mode
Define output behavior in alarm condition.
• Actual value
• Defined value
• 0 Hz
0 Hz
Failure frequency
Enter frequency output value in alarm
condition.
0.0 to 12
500.0 Hz
0.0 Hz
Invert output signal
Invert the output signal.
• No
• Yes
No
Switch output
Navigation
"Setup" menu → Pulse/frequency/switch output
Structure of the submenu for the switch output
Pulse/frequency/switch output
→
Operating mode
Switch output function
Assign diagnostic behavior
Assign limit
Assign flow direction check
Assign status
Switch-on value
Switch-off value
Failure mode
Invert output signal