Modbus RS485 Register Information
Proline Cubemass 300 Modbus RS485
274
Hauser
Navigation: Expert → Output → Double pulse out
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Pulse output (0987)
7041 to 7042
Float
Read
Positive floating-point number
Invert outp.sig. (0993)
6101
Integer
Read / Write
0 = Yes
1 = No
*
Visibility depends on order options or device settings
6.3.6
"Communication" submenu
"Modbus config." submenu
Navigation: Expert → Communication → Modbus config.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Bus address (7112)
4910
Integer
Read / Write
1 to 247
Baudrate (7111)
4912
Integer
Read / Write
0 = 1200 BAUD
1 = 2400 BAUD
2 = 4800 BAUD
3 = 9600 BAUD
4 = 19200 BAUD
5 = 38400 BAUD
6 = 57600 BAUD
7 = 115200 BAUD
Data trans. mode (7115)
4913
Integer
Read / Write
0 = RTU
1 = ASCII
Parity (7122)
4914
Integer
Read / Write
0 = Even
1 = Odd
2 = None/2 stop bits
3 = None/1 stop bit
Byte order (7113)
4915
Integer
Read / Write
0 = 0-1-2-3
1 = 3-2-1-0
2 = 2-3-0-1
3 = 1-0-3-2
Telegram delay (7146)
4916 to 4917
Float
Read / Write
0 to 100 ms
Failure mode (7116)
4920
Integer
Read / Write
1 = Last valid value
255 = NaN value
Bus termination (7155)
5774
Integer
Read
0 = Off
1 = On
Fieldb.writ.acc. (7156)
6807
Integer
Read / Write
0 = Read + write
1 = Read only
"Modbus info" submenu
Navigation: Expert → Communication → Modbus info
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Device ID (7153)
2547
Integer
Read
4-digit hexadecimal number
Device revision (7154)
4481
Integer
Read
4-digit hexadecimal number
Содержание Proline Cubemass 300