Description of device parameters
Proline Cubemass 300 Modbus RS485
12
Hauser
1
00914-2
A0029414
1
Direct access code
Note the following when entering the direct access code:
• The leading zeros in the direct access code do not have to be entered.
Example: Enter
"914"
instead of
"00914"
• If no channel number is entered, channel 1 is accessed automatically.
Example: Enter
00914
→
Assign variable
parameter
• If a different channel is accessed: Enter the direct access code with the corresponding
channel number.
Example: Enter
00914-2
→
Assign variable
parameter
Locking status
Navigation
Expert → Locking status (0004)
Description
Displays the active write protection.
User interface
• Hardware locked
• CT act.-all par.
• Temp. locked
Additional information
User interface
If two or more types of write protection are active, the write protection with the highest
priority is shown on the local display. In the operating tool all active types of write
protection are displayed.
Detailed information on access authorization is provided in the "User roles and
associated access authorization" and "Operating concept" sections of the Operations
Instructions for the device → 7
Selection
Options
Description
None
The access status displayed in the
Access status
parameter (→ 13) applies .
Only appears on local display.
Hardware locked
(priority 1)
The DIP switch for hardware locking is activated on the PCB board. This locks write
access to the parameters (e.g. via local display or operating tool) .
Temp. locked
(priority 3)
Write access to the parameters is temporarily locked on account of internal
processes running in the device (e.g. data upload/download, reset etc.). Once the
internal processing has been completed, the parameters can be changed once
again.