Electrical connection
Proline 200 FOUNDATION Fieldbus
18
Hauser
Removing a cable
2
1
3 4
3 (0.12)
A0032240
‣
To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between
the two terminal holes while simultaneously pulling the cable end out of the terminal.
5.2.2
Ensuring potential equalization
Requirements
No special measures for potential equalization are required.
For devices intended for use in hazardous locations, please observe the guidelines in the
Ex documentation (XA).
5.3
Ensuring the degree of protection
The measuring device fulfills all the requirements for the IP66/67 degree of protection, Type
4X enclosure.
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following steps
after the electrical connection:
1. Check that the housing seals are clean and fitted correctly.
2. Dry, clean or replace the seals if necessary.
3. Tighten all housing screws and screw covers.
4. Firmly tighten the cable glands.