Oxymax H COS21D
Technical data
Hauser
31
Ambient temperature
–5 to +135 °C (23 to 175 °F)
Ingress protection
IP 68 (10 m (33 ft) water column at 25 °C (77 °F) during 45 days, 1 mol/l
KCl)
10.3 Process
Process temperature
• COS21D-A and COS21D-C:
-5 to 135 °C (23 to 275 °F)
• COS21D-B:
-5 to 100 °C (23 to 212 °F)
Process pressure
• COS21D-A:
0 to 4 bar (0 to 58 psi)
• COS21D-B and COS21D-C:
0 to 12 bar (0 to 174 psi)
10.4 Performance characteristics
Response time
From air to nitrogen at 25 °C (77 °F)
• t
90
: < 30 s
• t
98
: < 60 s
Reference operating
conditions
Current value at air
• COS21D-A:
60 nA (40 to 80 nA)
• COS21D-B and COS 21D-C:
300 nA (180 to 500 nA)
Zero current
< 0.1 % of the current in air
Measured value
resolution
• COS21D-A:
10
μ
g/l (10 ppb)
• COS21D-B and COS21D-C:
1
μ
g/l (1 ppb)
Maximum measured
error
±1 % of measured value
1)
Reference temperature:
25 °C (77 °F)
Reference pressure:
1013 hPa (15 psi)
Reference application:
Municipal wastewater treatment