Wiring
Oxymax H COS21D
16
Hauser
4.3
Connection via junction box
To extend the sensor cable, the connection has to take place using a junction box RM (-->
Accessories). The extension to the transmitter is established by means of the CYK81 special
measuring cable.
The maximum permissible cable length is 100 m (328 ft).
4.4
Post-connection check
a0005724
Fig. 10: Wiring diagram with junction box RM
1
Sensor
2
Junction box
3
Extension cable
4
Transmitter
Instrument status and specifications
Remarks
Are the sensor, assembly, junction box or cable damaged?
Visual inspection
Electrical connection
Remarks
Are the installed cables strain-relieved and not twisted ?
Long enough length of cable core stripped and correct in terminal?
Check seating (pull slightly)
Are all the screws terminals properly tightened ?
Tighten
Are all the cable entries installed, tightened and sealed ?
For cable entries lateral: cable loops downwards
for water to be able to drip off.
Are all the cable entries installed downwards or lateral ?
YE
YE
BN
GN
GN
YE
WH
187
188
197
198
1
2
3
4
GN
BN
WH
BN
WH
GY
GY