Oxymax H COS21D
Installation
Hauser
11
3.4
Installation examples
3.4.1
Fixed installation (CPA442)
The CPA442 fixed installation assembly allows a sensor to be easily adapted to almost any
process connections from Ingold nozzles to Varivent or Triclamp connections.
This type of installation is very suitable for tanks and larger pipes. The sensor achieves a defined
immersion depth in the medium in an easy manner
3.4.2
Flow assembly (CPA240)
The CPA240 flow assembly offers up to three installation slots for sensors with a shaft diameter
of 12 mm (0.47"), a shaft length of 120 mm (4.72") and a process connection Pg 13.5.
It is very suitable for use in pipes or at hose connections.
During trace measurement in particular, ensure the assembly is completely vented to avoid
measuring errors.
a0005720
Fig. 4: CPA240 flow assembly with protective hood
a0005721
Fig. 5: Bypass installation
1
Main line
2
Medium removal
3, 6
Manually actuated or solenoid valves
4
Sampling
5
Flow assembly with installed sensor
7
Medium return