Installation
Oxymax H COS21D
12
Hauser
3.4.3
Retractable assembly (CPA475 or CPA450)
The assembly is designed for installation on tanks and pipes. Suitable nozzles must be available
for this.
Install the assembly at places with constant flow. The minimum pipe diameter is DN 80.
a0005722-en
Fig. 6: Suitable and unsuitable installation positions with a retractable assembly
1
Ascending pipe, best position
2
Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming
3
Horizontal pipe, installation with permissible installation angle (acc. to sensor version)
4
Overhead installation, impermissible due to missing electrolyte contact of the sensor electrodes
5
Down pipe, impermissible
!
Note!
Do not install the assembly at places where air cushions or foam bubbles can form.
Measuring errors can occur, if:
• the sensor is not immersed into the medium
• suspended particles are settled on the sensor membrane
• the sensor is installed overhead.
3.5
Post-installation check
• Sensor and cable undamaged?
• Cap undamaged?
• Compliance with permissible sensor installation position?
• Is the sensor installed in an assembly and is not suspended from the cable?
• Avoid moisture by rain by putting the protective cap on the assembly?