Installation
Oxymax COS61D
12
Hauser
5
Installation
5.1
Mounting conditions
5.1.1
Orientation
45°
135°
45°
0°
180°
A0032281
2
Installation angle
Recommended installation angle
Possible installation angle
Inadmissible installation angle
The sensor must be installed at an angle of inclination in an assembly, holder or
appropriate process connection. Recommended angle: 45° to prevent the attachment of air
bubbles. At angles of inclination of 45 to 135°, air bubbles at the oxygen-sensitive
membrane may increase the measured value.
The sensor can be installed up to the horizontal in an assembly, holder or suitable process
connection. The optimum installation angle is 45°.
Other angles and upside-down installation are not recommended. Reason: Possible
sediment formation and resulting falsification of measured value.
Follow the instructions for installing sensors in the Operating Instructions for the
assembly used.
5.1.2
Mounting location
1. Choose a mounting location that is easy to access.
2. Ensure that upright posts and assemblies are fully secured and vibration-free.
3. Choose a mounting location with an oxygen concentration that is typical for the
application.