MMP20 SONO-DIS (option D)
Hauser
Page 45
Note:
Only a maximum of 6 values can be saved in the list. Older values are removed from
the list and are no longer used to calculate the average.
To delete the measurement series, press the "Down" key
.
TIP:
14.3.3
"Water calculation" operating mode
Only the moisture is measured in this mode. A mean value is calculated from up to six individual
values. The water quantity per m³ is calculated from this mean. The density of the material being
measured must be set to ensure the value is calculated correctly.
To delete the measurement series, press the "Down" key
.
To set the density, press the "Up" key in the measurement window
.
The following screen then appears:
Set the density of your material with the "Up"
and
"Down"
keys and press the "Measure" key
to confirm
your entries.
You return automatically to the measurement window.
You can exit the density setting with the "Settings" key
without changing the old density value.
Individual mea-
surement values
Number of the selected
calibration
Remaining bat-
tery
Average water con-
tent in l/m³
Density in kg/dm³
With only six measurements in the "mean value" operating
mode, you can obtain a representative measurement result
from all measuring points over a large quantity of material.