Micropilot M FMR24x/250, new software version “Pulse Master eXact” 01.05.00

The new SW version Pulse Master eXact 01.05.00 has been introduced in CW39/2006 for the following instruments:

• FMR240

HART and PROFIBUS PA

• FMR244

HART and PROFIBUS PA

• FMR245

HART and PROFIBUS PA

• FMR250

HART and PROFIBUS PA

The following changes have been made to the software function:

• Introduction of new SW parameter for differentiation between liquids and solids applications
• Filter settings of new SW with application parameters

The new software is compatible with these operating tools:

• ToF Tool Fieldtool Package as of version 4.00
• HART Communicator DXR375, revision1, DD revision 1
• Fieldcare version 2.02.02 / DTM library release C-2006

A flash update from software version 01.04.00 to 01.05.00 is not possible due to different hardware. A hardware exchange is required in
any case.

• in case of faulty electronics, the electronics module must be exchanged with the same type with SW 4.0
• if the HF module is faulty, it must be exchanged with the same type for SW4.0
• if an upgrade to 5.0 is desired (for advanced dynamics), HF module and electronics must always be exchanged together

• in case of faulty electronics, the electronics module can only be exchanged with the same type with SW5.0
• if the HF module is faulty, it can only be exchanged with the same type for SW5.0

• spare electronics FMR24x/250 HART software version 01.05.00

order number

7102 6754

• spare electronics FMR24x/250 PROFIBUS PA SW01.05.00

order number

7102 6819

• spare electronics FMR24x/250 FOUNDATION Fieldbus SW01.05.00

order number

7102 6820 (from 1. quarter 07)

• spare HF module advanced dynamics, 26 GHz, for SW 01.05.00

order number

7102 6572

The conversion is possible for non-Ex instruments or Ex ia / Ex d instruments, however not for solids instruments with certification
WHG / SIL or FM.
Together with the new software version, we offer 2 new HF modules with enhanced dynamics for liquids applications:

• spare HF module max. measuring range = 40 m

order number

71026572

• spare HF module max. measuring range = 70 m

order number

5202 4953

A HF module can only be changed in combination with a change to electronics SW5.0.

Hints for repair and spare parts for FMR24x

Instruments with software version Pulse Master 01.04.00:

Instruments with software version Pulse Master eXact 01.05.00:

Conversion of an instrument FMR24x for special applications (advanced dynamics)

Service Information
PC-Maulburg

Hauser Maulburg

page 1 of 8

Contact person:

Hans-Joachim Freytag (fr)
Service MSR
Telephone: (+49) 07622 28-2541

[email protected]

Classification:

for information
important, handle re-actively
urgent, handle actively
safety relevant

ID:

Pages:
Date:

X

SIF8901301en0

8
Oct. 2006

please turn over

Содержание Micropilot M FMR244

Страница 1: ... HF module is faulty it can only be exchanged with the same type for SW5 0 spare electronics FMR24x 250 HART software version 01 05 00 order number 7102 6754 spare electronics FMR24x 250 PROFIBUS PA SW01 05 00 order number 7102 6819 spare electronics FMR24x 250 FOUNDATION Fieldbus SW01 05 00 order number 7102 6820 from 1 quarter 07 spare HF module advanced dynamics 26 GHz for SW 01 05 00 order num...

Страница 2: ...tory map valid matrix position 0D34 horn antenna p ic a arabol antenn Enter the mapping range must be free of obstructions without factory map error W511 In the operating matrix enter the blocking distance mm inch at group selection extended calibration sub menu calibration selection general field blocking distance matrix position 059 In the operating matrix enter the mapping range max range is 17...

Страница 3: ... level of the low gain peak is used as a reference This results in an average noise level raised by about 20 dB approx 70 dB This noise level cannot be used by itself for envelope curve evaluation in the field The evaluation is based on the dynamics between echo level and noise level dynamic approx 60 dB evaluation approx 90 dB Measuring range change via reset Depending on the instrument version 3...

Страница 4: ...ot active 1 active fact map valid 0 not active 1 active max ampl over FAC envelope energy present edge parameter Mapping Edge First echo Tank bottom detection Module Antenna table present amplitude amplitude over mapping amplitude over FAC unfiltered distance present edge parameter present first echo factor device name order code 0D0 0D1 0D2 0D3 0D4 0D5 0D6 0D7 0D8 0D9 0DC 0DD back to function gro...

Страница 5: ...isplay module output on alarm xx xx mA output on alarm max scale diameter vessel m tank shape dome ceiling horizontal cyl flat ceiling sphere bypass stilling well customer unit see table customer unit see table release Oct 06 jump to matrix field 008 too big unknown linearisation linear manual medium propert DK unknown DK 1 6 1 9 DK 1 9 2 5 DK 2 5 4 0 DK 4 0 7 0 DK 7 0 process conditions standard ...

Страница 6: ...only Micropilot M FMR24x 250 operating structure HART display module software version 01 05 00 language back to home 3 9999 seconds format display decimal 1 16 display test not for ToF Tool off on 1s remote control off on reset separat character point komma No of decim serial No distance unit m ft mm inch English Deutsch Francais Espanol Italiano Nederlands Japanese 100 r w customer matrix 200 rea...

Страница 7: ...large reflectivity 0 7 9 process condition Standard 2 3 160 0 15 4 3 10 10 Calm Surface 3 4 120 0 32 7 10 10 10 Turbulent Surface 2 4 200 0 25 5 5 10 10 Add Agitator 2 4 200 0 25 7 5 10 10 Fast change 2 2 160 0 5 1 0 10 10 Test no filter 2 2 160 0 1 0 0 10 10 Solid 1000 200 tank shape unknown 15 72 0 metal silo 15 72 0 concrete silo 15 72 0 bin bunker 15 72 0 dome 15 72 0 stockpile 15 72 0 conveye...

Страница 8: ...flectivity process condition Standard 0 10 Calm Surface 0 10 Turbulent Surface 1 10 Add Agitator 1 10 Fast change 0 0 Test no filter 0 0 Solid 60 600 0 2000 1 3000 1500 tank shape unknown 0 0 0 metal silo 0 0 0 concrete silo 0 0 0 bin bunker 0 0 0 dome 0 0 0 stockpile 0 0 0 conveyer belt 0 0 0 medium condition unknown 1 6 1 9 very small reflectivity 1 9 2 5 small reflectivity 2 5 4 middle reflecti...

Отзывы: