manualshive.com logo in svg
background image

Safety instructions

Micropilot FMR51, FMR57

2

Safety instructions

1.1 

Designated use

The horn protection is a retrofitting part for Micropilot FMR51 and FMR57 level

measurement devices.

1.2 

Installation and commissioning

1.2.1 

Authorised personnel

All operations described in this mounting instructions manual must be carried out only

by trained specialist personnel authorised by the operator. During work on and with the

device the required personal protection equipment must always be worn.

1.2.2 

Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specific

hazards, e.g. vessel overfill or damage to system components through incorrect

mounting or adjustment.

1.3 

Operational safety

1.3.1 

General safety instructions

The safety information in the operating instructions of the respective sensor must be

observed.

1.3.2 

Hazardous areas

• in gas explosion hazardous areas

• in dust explosion hazardous areas (apparatus group IIIC)

NOTICE

Danger of explosion

In gas explosion hazardous areas: Avoid electrostatic charging (e.g. through friction,

cleaning, maintenance, strong medium flow).

Содержание Micropilot FMR57

Страница 1: ...Products Solutions Services Special documentation Micropilot FMR51 FMR57 Horn protection For 80 mm 3 and 100 mm 4 horn antenna SD01084F 00 EN 01 12 71212877 ...

Страница 2: ...ion equipment must always be worn 1 2 2 Warning about misuse Inappropriate or incorrect use of the instrument can give rise to application specific hazards e g vessel overfill or damage to system components through incorrect mounting or adjustment 1 3 Operational safety 1 3 1 General safety instructions The safety information in the operating instructions of the respective sensor must be observed ...

Страница 3: ...2 Principle of operation The horn protection is suitable for Micropilot FMR51 FMR57 level measurement devices in applications with strong buildup and condensation inside the horn antenna Thanks to seals on the horn protection and antenna socket moisture penetration can be prevented The contact ring provides an electrical connection between holding ring and horn antenna When using the horn protecti...

Страница 4: ...ction 8 Screws horn protection 1 Loosen the screws 3 on the antenna socket and remove horn antenne 2 Shift O ring seal 6 contack ring 5 and holding ring 4 onto the antenna cone 3 Plug horn protection 7 onto the antenna opening and direct the holes to the holding ring 4 4 Tighten the screws 8 crosswise until the horn protection 7 and holding ring 4 are assembled 5 Insert flat seal 1 between horn an...

Страница 5: ...eal Silicone Flat seal PTFE 4 1 2 Process conditions Vessel pressure max 0 5 bar 7 252 psi Process temperature max 130 C 266 F 4 2 Dimensions 4 2 1 Horn protection for horn antenna 80 mm 3 in 1 237 9 33 DN100 mm in ø96 3 78 ød A0019143 2 Horn protection for horn antenna 80 mm 3 1 Associated version of FMR51 FMR57 feature 070 Antenna option BC Horn 80mm 3in ...

Страница 6: ...mm 4 in 2 450 17 72 ø116 4 57 mm in ød DN150 A0019144 3 Horn protection for horn antenna 100 mm 4 4 3 Ordering information Antenna size Order code of horn protection 80mm 3 71105890 100mm 4 71105889 2 Associated version of FMR51 FMR57 feature 070 Antenna option BD Horn 100mm 4in ...

Страница 7: ......

Страница 8: ...www addresses endress com 71212877 71212877 ...

Отзывы:

Похожие инструкции для Micropilot FMR57