Memosens COS51E

Technical data

Hauser

43

Maximum measured
error

 2)

COS51E-****TN (black membrane cap):

≤ ±1 % of the measured value

COS51E-****TF (white membrane cap):

≤ ±1 % of the measured value

Limit of detection (LOD)

 3)

COS51E-****TN (black membrane cap):

10 ppb

COS51E-****TF (white membrane cap):

5 ppb

Limit of quantification
(LOQ)

 3)

COS51E-****TN (black membrane cap):

20 ppb

COS51E-****TF (white membrane cap):

10 ppb

Repeatability

COS51E-****TN (black membrane cap):

20 ppb

COS51E-****TF (white membrane cap):

100 ppb

Long-term drift

 4)

Zero-point drift:

< 0.1 % per week

Measuring range drift:

< 0.1 % per week

Polarization time

< 60 minutes

Intrinsic oxygen

consumption

• COS51E-****TN:

approx. 90 ng/h in air at 25 °C (77 °F)

• COS51E-****TF:

approx. 270 ng/h in air at 25 °C (77 °F)

Electrolyte

Alkaline saline solution

12.4   Environment

Ambient temperature

range

-5 °C ≤ T

a

 ≤ 60 °C (T6)

23 °F ≤ T

a

 ≤ 140 °F (T6)

Storage temperature range

• Filled with electrolyte:

–5 to 60 °C (20 to 140 °F)

• Without electrolyte:

–20 to 60 °C (0 to 140 °F)

Degree of protection

IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 30 days)

12.5   Process

Process temperature range

-5 ≤ T

p

 ≤ 60 °C (T6)

41 °F ≤ T

p

 ≤ 140 °F (T6)

2)

In accordance with IEC 60746-1 at rated operating conditions

3)

According to DIN EN ISO 15839. The measured error contains all the uncertainties of the sensor and transmitter (measuring chain). It does not
contain all the uncertainties caused by the reference material and adjustments that may have been performed.

4)

Under constant conditions

Содержание Memosens COS51E

Страница 1: ...Products Solutions Services Operating Instructions Memosens COS51E Amperometric oxygen sensor with Memosens 2 0 technology BA02146C 07 EN 02 22 00 71556090 2022 01 19...

Страница 2: ......

Страница 3: ...nting the sensor 14 5 3 Installation examples 16 5 4 Post mounting check 21 6 Electrical connection 22 6 1 Connecting the sensor 22 6 2 Ensuring the degree of protection 22 6 3 Post connection check 2...

Страница 4: ...If necessary Consequences of non compliance if applicable Corrective action This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or more serious injurie...

Страница 5: ...on for the relevant sensor Operating Instructions for the transmitter used Operating Instructions for the cable used Safety data sheet for the relevant electrolyte solutions In addition to these Opera...

Страница 6: ...ludge basins Monitoring the oxygen content in the wastewater treatment plant outlet Monitoring measuring and regulating the oxygen content in public waters and fish farming water Use of the device for...

Страница 7: ...aning system is connected switch it off before removing a sensor from the medium If you need to test the cleaning function while cleaning is in progress wear protective clothing goggles and gloves or...

Страница 8: ...Screw thread for protection guard 12 Sealing ring 13 Screw thread for membrane cap 14 Membrane cap 6 A0046014 3 Sensor head top view 6 Cathode The sensor consists of the following functional units Se...

Страница 9: ...electrode at a constant potential The advantages are a much higher signal accuracy and significantly extended calibration intervals 3 4 Membrane body The oxygen dissolved in the medium is transported...

Страница 10: ...calibration Possibility to reset to factory calibration In the case of sensors with replaceable measuring elements the number of calibrations per measuring element and for the entire sensor Operating...

Страница 11: ...l packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact your supplier or your local Sales Center 4 2 Product identific...

Страница 12: ...ngen 4 3 Scope of delivery The scope of delivery comprises Ordered version of the sensor with protective cap filled with tap water to protect the membrane Accessories set with the following contents 2...

Страница 13: ...Memosens COS51E Mounting Endress Hauser 13 5 Mounting 5 1 Mounting requirements 5 1 1 Dimensions 171 6 7 9 7 14 5 7 5 9 4 191 9 7 55 40 1 57 A0045976 4 Dimensions in mm inch...

Страница 14: ...or installing sensors in the Operating Instructions for the assembly used 5 1 3 Mounting location 1 Choose a mounting location that is easy to access 2 Ensure that upright posts and assemblies are ful...

Страница 15: ...t non grounded metallic assemblies may be live and as such are not safe to touch When using metallic assemblies and installation equipment national grounding provisions must be observed To fully insta...

Страница 16: ...and pipes This requires the availability of suitable process connections Install the assembly in a place with uniform flow conditions The pipe diameter must be at least DN 80 5 3 4 1 2 3 A A 2 4 A004...

Страница 17: ...exdip CYA112 1 2 3 6 4 5 7 A0042858 9 Chain holder on upright post 1 Weather protection cover CYY101 2 Transmitter 3 Chain 4 Wastewater assembly Flexdip CYA112 5 Oxygen sensor 6 Basin rim 7 Holder Fle...

Страница 18: ...2159 12 Dimensions in mm inch 1 Cable run with strain relief and rain shield 2 Fixing ring for rope and chains with terminal screw 3 Eyelets 15 3 x 120 for anchoring 4 Plastic float resistant to salt...

Страница 19: ...solenoid valves 1 Main pipe 2 Medium return 3 Oxygen sensor 4 7 Manually actuated or solenoid valves 5 Flow assembly COA250 A 6 90 pipe elbow 8 Medium removal 5 3 4 Universal flow assembly Flowfit CYA...

Страница 20: ...m return Mount the sensor in the assembly in accordance with the Operating Instructions BA00495C There must be a minimum flow rate of 100 ml h 0 026 gal h Take increased response times into account As...

Страница 21: ...g Endress Hauser 21 5 4 Post mounting check 1 Are the sensor and cable undamaged 2 Is the orientation correct 3 Is the sensor installed in an assembly and is not suspended from the cable 4 Avoid the p...

Страница 22: ...d Exercise care when carrying out the work Otherwise the individual types of protection Ingress Protection IP electrical safety EMC interference immunity agreed for this product can no longer be guara...

Страница 23: ...ular intervals to ensure reliable measurement Operating Instructions for Memosens COS51E BA02146C Operating Instructions for the transmitter used such as BA01245C if using the Liquiline CM44x or Liqui...

Страница 24: ...is sufficient calibration of the sensor slope When changing from process to calibration conditions a longer polarization time and an adjustment of the temperature to the environment must be taken int...

Страница 25: ...ient effects direct sunlight drafts during this time 4 Is the measured value display on the transmitter stable Perform the calibration in accordance with the Operating Instructions for the transmitter...

Страница 26: ...K 0 0 1 000 550 1800 0 938 1050 3450 0 885 1550 5090 0 834 50 160 0 994 600 1980 0 932 1100 3610 0 879 1600 5250 0 830 100 330 0 988 650 2130 0 927 1150 3770 0 874 1650 5410 0 825 150 490 0 982 700 23...

Страница 27: ...ve sea level current air pressure 1009 hPa S 9 45 mg l K 0 943 L 0 996 M 1 02 Calibration value C 9 05 mg l Factor K in the table is not required if the measuring device returns the absolute air press...

Страница 28: ...reate medium flow No electrolyte in the measuring chamber Refill or replace electrolyte Buildup on the membrane cap Clean sensor carefully Displayed value too high Polarization ended Wait for polariza...

Страница 29: ...eshooting instructions in the Operating Instructions for the transmitter Check the transmitter if necessary If one of the above mentioned faults occurs 1 Unscrew membrane cap from sensor 2 Dry the ele...

Страница 30: ...heric pressure if not already the same If the process pressure and atmospheric pressure are the same this step is not required 3 After 10 minutes measure the oxygen saturation index in air Decide usin...

Страница 31: ...ith copious amounts of clean water Use a fully automated cleaning system for regular automated cleaning 9 2 2 Cleaning the cathode The cathode only needs to be cleaned if it is coated or silver plated...

Страница 32: ...the green polishing foil Hold the sensor as vertically as possible Move the cathode in small circles over the polishing foil 4 A0046399 5 A0046401 4 Clean the cathode surface with the pink polishing f...

Страница 33: ...Memosens COS51E Maintenance Endress Hauser 33 6 Mount the sensor 37...

Страница 34: ...safe professional and swift product returns please contact your local Sales Center for information on the procedure to be followed and general conditions 10 3 Spare parts and consumables Parts of the...

Страница 35: ...enses rinse eyes with water for a few minutes and contact a doctor In case of skin contact Take off wet clothing immediately wash the skin or take a shower Also pay attention to the safety datasheet o...

Страница 36: ...res The electrolyte depletion can be logged using a suitable transmitter By setting warning limits it is possible to properly schedule sensor maintenance Theoretical operating time at pO2 210 mbar and...

Страница 37: ...In case of skin contact Take off wet clothing immediately wash the skin or take a shower 1 A0046403 2 A0046402 3 A0046414 1 Fill the membrane cap halfway with fresh electrolyte Remove any air bubbles...

Страница 38: ...just the process pressure at the transmitter if it differs from the atmospheric pressure otherwise a comparison will not be possible 4 After about 10 minutes measure the oxygen saturation index in air...

Страница 39: ...www endress com coa451 Technical Information TI00368C Flexdip CYH112 Modular holder system for sensors and assemblies in open basins channels and tanks For Flexdip CYA112 water and wastewater assembli...

Страница 40: ...ct page www endress com CYR10B Technical Information TI01531C 11 1 4 Transmitter Liquiline CM44 Modular multi channel transmitter for hazardous and non hazardous areas HART PROFIBUS Modbus or EtherNet...

Страница 41: ...d documentation of sensor management Sensor calibrations stored in database Product Configurator on the product page www endress com cyz71d Technical Information TI00502C 11 1 5 Maintenance kit COV45...

Страница 42: ...19 Measuring cable CYK10 12 3 Performance characteristics Response time 1 At 20 C 68 F COS51E TN black membrane cap for standard response time t90 3 minutes t98 8 minutes COS51E TF white membrane cap...

Страница 43: ...ption COS51E TN approx 90 ng h in air at 25 C 77 F COS51E TF approx 270 ng h in air at 25 C 77 F Electrolyte Alkaline saline solution 12 4 Environment Ambient temperature range 5 C Ta 60 C T6 23 F Ta...

Страница 44: ...6 7 9 7 14 5 7 5 9 4 191 9 7 55 40 1 57 A0045976 20 Dimensions in mm inch Weight 0 3 kg 0 7 lbs Materials Parts in contact with medium Sensor shaft POM Membrane cap POM Working electrode Gold Counter...

Страница 45: ...Memosens COS51E Technical data Endress Hauser 45 Membrane thickness COS51D 0 Approx 50 m COS51D 1 Approx 25 m...

Страница 46: ...or 43 Measured variables 42 Measurement function 38 Measuring cable 39 Measuring principle 9 Measuring range 42 Measuring system 14 Mechanical construction 44 Mounting 13 Mounting requirements 13 N Na...

Страница 47: ......

Страница 48: ...www addresses endress com 71556090 71556090...

Отзывы: