Mounting
Memosens COS51E
20
Hauser
1
3
5
6
2
7
2
2
4
A0032920
16
Connection diagram
1
Main pipe
5
Medium inflow
2
Manually actuated or solenoid valves
6
Flow assembly
3
Medium outlet
7
Medium removal
4
Medium return
Mount the sensor in the assembly in accordance with the Operating Instructions
(BA00495C).
There must be a minimum flow rate of 100 ml/h (0.026 gal/h).
‣
Take increased response times into account.
As an alternative to bypass operation, direct the sample flow from a filter unit with an
open outlet through the assembly:
2
1
3
4
A0032921
17
Flow assembly with open outlet
1
Pump
2
Assembly
3
Open outlet
3
Filter unit