Principle of factor calibration
x
Measured value
y
Target sample value
a
Factory calibration
b
Factor calibration
Offset
With the "Offset" function, the measured values are offset by a constant amount (added or
subtracted).
1
2
3
4
5
6
7
8
1
2
3
4
5
6
x
y
a
b
A0039330
28
Principle of an offset
x
Measured value
y
Target sample value
a
Factory calibration
b
Offset calibration
8.1.3
Signal filter
The sensor is fitted with an internal signal filter function in order to adapt the
measurement flexibly to different measuring requirements. Fluorescence measurements
may have a low signal-to-noise ratio. In addition, there may be disturbances from air
bubbles or contamination for example.