Electrical connection
Memosens CFS51
22
Hauser
6
Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the sensor
The following connection options are available:
• Via M12 plug (version: fixed cable, M12 plug)
• Via the cable of the device to the plug-in terminals of the input of the transmitter
(version: fixed cable, wire end ferrules)
85
86
85
2DS
1
2
86
97
88
87
98
97
88
87
98
Sensor 1
Sensor 2
PK
GY
GN
YE
85
86
85
1
2
86
97
88
87
98
Sensor 1
PK
GY
GN
YE
BN
WH
A0042911
18
Device connection to input (left) or with M12 plug (right)
The device is available with the following fixed cable lengths:
• 3 m (9.84 ft)
• 7 m (22.97 ft)
• 15 m (49.22 ft)
6.1.1
Connecting the cable shield
The device cable must be shielded cables.
Only use terminated original cables where possible.