background image

KA457C/07/EN/13.10
71128660

Brief Operating Instructions

Memograph CVM40

Graphic transmitter for inline photometers and data 
manager

Содержание Memograph CVM40

Страница 1: ...KA457C 07 EN 13 10 71128660 Brief Operating Instructions Memograph CVM40 Graphic transmitter for inline photometers and data manager...

Страница 2: ......

Страница 3: ...symbols 5 1 6 Document symbols 5 2 Identification 6 2 1 Scope of delivery 6 3 Installation 7 3 1 Measuring system 7 3 2 Installation conditions 8 3 3 Installation instructions 9 4 Wiring 11 4 1 Electr...

Страница 4: ...erating Instructions and must adhere to them Before commissioning the entire measuring point check all the connections Ensure that electrical cables and hose connections are not damaged Do not operate...

Страница 5: ...ymbols Warning This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if ignored Caution This symbol alerts you to possible faults which could arise from incor...

Страница 6: ...per your order 2 fastening clips USB interface cable length 1 5 m 4 9 ft Optional secure digital SD card Link to PC operating and configuration software on CD ROM Lamp voltage sense tool Brief Operati...

Страница 7: ...stem A complete measuring system comprises Memograph CVM40 An optical sensor e g OUSAF44 Flow assembly e g OUA260 Cable set e g OUK40 a0012117 Fig 1 Example of a measuring system 1 Cable set OUK40 2 M...

Страница 8: ...ondensation Caution To avoid heat accumulation please always ensure that the unit is sufficiently cooled Maintain distance from strong magnetic fields see Operating Instructions on CD ROM Section 10 T...

Страница 9: ...ng the device Panel cutout and installation design dimensions a0012075 Fig 2 Dimensions panel cutout 29 5 1 16 76 2 99 199 2 7 84 216 03 8 51 137 5 5 41 29 1 14 39 07 1 54 136 5 35 136 5 35 195 23 7 6...

Страница 10: ...from the walls and other units 2 Hold the unit horizontally and then hang the two jack screws in the openings opposite either on left or right housing side or at top or bottom 3 Tighten the screws on...

Страница 11: ...ection must be made before all other connections Any interruption in the ground can cause danger Before commissioning compare the supply voltage with the information specified on the nameplate left ha...

Страница 12: ...e decrease Turn counter clockwise voltage increase 2 V1 1 Lamp voltage V1 3 Lamp sense V1 4 Lamp sense V1 2 Lamp voltage 3 S1 1 Photo diode anode For measurement detector S1 2 Photo diode cathode For...

Страница 13: ...pe to the transmitter Failure in doing so can damage the lamp By using the lamp voltage sense tool the lamp voltage can be adjusted approximately to the rated value before connecting the sensor Connec...

Страница 14: ...ph CVM40 14 Endress Hauser a0012068 Fig 5 Wiring diagram 1 Power supply 6 Thermocouples TC 2 Binary inputs D 7 Voltage U 3 Analog outputs O 8 Current I 4 Analog inputs 9 Resistance thermometer RTD 5 P...

Страница 15: ...his makes the connection very quick and easy The spring terminals are actuated with a slotted screwdriver size 0 Note No ferrules need to be used when connecting flexible wires to spring terminals 4 4...

Страница 16: ...a complete measuring system the unit is factory calibrated and preset with the related sensor and armature Operation of measuring loop based on basic settings is guaranteed when switching on the devi...

Страница 17: ...the Display mode alternating display indicating what percentage of the SD card or USB stick has already been written to Status symbols are displayed for the following functions alternate with the memo...

Страница 18: ...fault setting for the operating language is English A different operating language can be configured in the main menu Press the Navigator select Language a0012193 5 3 Setup 5 3 1 General information N...

Страница 19: ...ove the corresponding keys on the screen Free fields mean that no function is currently assigned to the keys in question Press the navigator the main menu is displayed Using the navigator select the S...

Страница 20: ...te time communication safety etc Inputs universal inputs signal type input range etc Inputs digital inputs function designation etc Inputs for mathematics channels Outputs relay or analog pulse output...

Страница 21: ...ndress Hauser 21 5 3 3 Setup Inputs Note Depending on the selected function the unit s user interface adapts itself so that each time only parameters that are required for safe unit functioning have t...

Страница 22: ...ical sensors Start Select sensor type Enter optical pathlength Enter the channel name Enter attached sensor information order code and serial number Select the plot type Enter the number of decimal pl...

Страница 23: ...sensor type OUSAF44 OUSAF45 etc The channel is switched off if no sensor is selected factory default Picklist switched off OUSAF11 AF12 AF13 DP11 OUSAF21 OUSAF22 OUSAF23 OUSAF44 OUSAF45 OUSAF46 OUSTF...

Страница 24: ...one X Y Zoom start If the total transmitter range is not being used then the lower value of the required range can be entered here higher resolution Example transmitter 0 to 2 5 AU required range 0 7...

Страница 25: ...vidual channel output OUSAF21 22 23 only Enable the absorbance measurement on individual channels of a dual beam optical sensor if required Absorbance monitoring OUSTF10 OUSAF30 only Enable absorbance...

Страница 26: ...conditions 8 Installation instructions 9 Instrument configuration 19 Interface connection 15 K Key assignment 16 L Lamp voltage adjustment 13 Lamp voltage sense tool 13 Language 18 M Measuring system...

Страница 27: ...ning Returned part cleaned with explosive oxidising dangerous for the environment biological risk radioactive Please tick should one of the above be applicable include safety data sheet and if necessa...

Страница 28: ...KA457C 07 EN 13 10 FM SGML 6 0 DT...

Отзывы: