Liquistation CSF39
Electrical connection
Hauser
39
6.9
Post-connection check
L
WARNING
Connection errors
The safety of people and of the measuring point is at risk! The manufacturer does not
accept any responsibility for errors that result from failure to comply with the instructions
in this manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Instrument status and specifications
‣
Are the device and all the cables free from damage on the outside?
Electrical connection
‣
Are the mounted cables strain relieved?
‣
Are the cables routed without loops and cross-overs?
‣
Are the signal cables correctly connected as per the wiring diagram?
‣
Are all plug-in terminals securely engaged?
‣
Are all the connection wires securely positioned in the cable terminals?