Technical data
Liquistation CSF39
184
Hauser
15.12 Process
Medium temperature range
2 to 50 °C (36 to 122 °F)
Process pressure
• Unpressurized, open channel (unpressurized sampling)
• Max. 0.8 bar piping (only with shutoff/inlet valve)
Medium properties
Vacuum pump
Capacitance level measurement used for:
• Sample media has to be free of abrasive substances.
• Media that tend to create a lot of foam or contain fats and grease
• Media with a conductivity < 30 µS/cm
Peristaltic pump
Sample media has to be free of abrasive substances.
Pay attention to the material compatibility of the wetted parts.
Process connection
•
Vacuum pump:
Suction line ID 10 mm (3/8"), 13 mm (1/2"), 16 mm (5/8") or 19 mm (3/4")
•
Peristaltic pump:
Intake hose ID 10 mm (3/8")
15.13 Mechanical construction
Dimensions
See the "Installation" section → 15
Weight
Sampler version
Weight
Plastic version without refrigeration
91 kg (201 lbs)
Plastic version with refrigeration
101 kg (223 lbs)
Stainless steel version with stand and refrigeration
146 kg (322 lbs)
Stainless steel version with refrigeration
125 kg (276 lbs)
Materials
Plastic polystyrene VO can change color when exposed to direct sunlight. For outdoor
use without a weather protection cover, the use of Plastic ASA+PC V0 is
recommended. The functionality is not affected by the discoloration.
Non-wetted parts
Cabinet housing
Plastic polystyrene V0
For standard applications in wastewater treatment plants and
environmental monitoring
Plastic ASA+PC V0
For industrial wastewater treatment plants with an aggressive
atmosphere
Stainless steel V2A (1.4301)
For standard applications in wastewater treatment plants and
environmental monitoring
Stainless steel V4A (1.4571)
For industrial wastewater treatment plants with an aggressive
atmosphere
Sample compartment inner lining
Plastic PP
Содержание Liquistation CSF39
Страница 2: ......
Страница 188: ...www addresses endress com ...