Liquistation CSF33
Diagnostics menu
Hauser
23
2.6
Term information
The following information is displayed:
•
Operating hours device:
Displays the total operating hours of the device in days, hours and minutes
•
Operating hours cooling
(only for the version with a climate control module):
Displays the total operating hours of the compressor in days, hours and minutes
•
Overfill sensor
(for version with vacuum pump):
Number of times a safety switchoff has been caused by conductivity 2
•
Dosing valve
(for version with vacuum pump):
Number of times the dosing valve is actuated; -> corresponds to the number of samples taken
•
Vacuum pump
(for version with vacuum pump):
Displays the pump operating time in hours and minutes
•
Sample totalizer
(for version with peristaltic pump):
Number of all samples taken and sample errors
•
Pump tube life
(for version with peristaltic pump):
Displays how old the tube is in days, hours and minutes
This counter must be reset when a tube is replaced.
Set the specific counter reading to zero with "Reset".
2.7
Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
• Relay contact opening or closing
Only current values are simulated. It is not possible to use the simulation function to calculate the
totalized value for the flow or rainfall.
Device Reset
Options
• OK
• ESC
Restart and keep all the settings
Factory default
Options
• OK
• ESC
Restart with factory settings
Settings that have not been saved are lost.
Power supply
Display
• Supply Voltage 1V2
• Supply Voltage 3V3
• Supply Voltage 12V5
• Supply Voltage 24V
• Temperature
Diagnostics/Systemtest/Reset
Function
Options
Info