Liquistation CSF33
Diagnostics menu
Hauser
13
2
Diagnostics menu
The Diagnostics menu contains all the information on the device status.
Furthermore, various service functions are available.
The following messages are directly displayed every time you enter the menu:
• "Most important message"
Diagnostics message recorded with the highest criticality level
• "Last message"
Last diagnostics message recorded
All the other functions in the Diagnostics menu are described in the following chapters.
Diagnostics messages associated with sampling are deleted under the following conditions:
• Diagnostics messages caused by sampling are deleted automatically with the next successful
sampling.
• Diagnostics messages caused by the level of medium in the bottle are deleted the next time the
program is started.
If diagnostics message "F313 Liquidsensor 1" appears five times in succession when running a
program, the active program is aborted for reasons of safety. This behavior on the part of the device
cannot be modified by deactivating the diagnostics message under Menu/Setup/General
settings/Diagnostics.
2.1
Diagnostics list
All the current diagnostics messages are listed here.
A time stamp is available for each message. Furthermore, the system also displays the configuration and
description of the message as saved in "Menu/Setup/General settings/Diagnostics/Device behavior".
For this purpose, select the appropriate message and press the navigator.