Technical data
Liquistation CSF28
66
Hauser
Relay outputs
Electrical specification
Relay type
2 x changeover contact, coupled with binary output
Maximum load
All other relays: 2.0 A
Relay switching capacity
Relay coupled with binary output
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cosΦ = 0.8 to 1
5 A
100,000
24 V DC, L/R = 0 to 1 ms
5 A
100,000
Minimum load (typical)
• min. 100 mA at 5 V DC
• min. 1 mA at 24 V DC
• min. 5 mA at 24 V AC
• min. 1 mA at 230 V AC
15.3 Protocol-specific data
Web server
The Web server enables full access to the device configuration, measured values,
diagnostic messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G
routers with a user-defined IP address.
TCP port
80
Supported features
• Remote-controlled device configuration
• Save/restore device configuration (via SD card)
• Logbook export (file formats: CSV, FDM)
• Access to Web server via DTM or Internet Explorer
15.4 Power supply
Supply voltage
100 to 120/200 to 240 V AC ±10 %, 50/60 Hz
Power consumption
• Version with vacuum pump: 290 VA
• Version with peristaltic pump: 290 VA
Electrical connection
See the "Electrical connection" section ()
Cable entries
1 x M25, 1 x M20 cable gland
Permitted cable diameter:
• M20x1.5 mm: 7 to 13 mm (0.28 to 0.51")
• M25x1.5 mm: 9 to 17 mm (0.20 to 0.67")
Mains fuse
T3.15A (for 230V power supply)