Diagnostics and troubleshooting
Liquistation CSF28
52
Hauser
Function
Info
Distribution arm
Only for bottle configurations with more than one bottle.
Test distribution arm
When the menu item is activated, the distribution arm undergoes a test run.
Afterwards, the system moves to each position in succession and the position is
displayed. In the case of plate distribution, the arm moves left and right to
ensure the bottles are numbered consecutively.
Calibrate the distribution arm if the arm is not positioned precisely over the
bottles.
Detailed list of power supply to instrument.
The actual values can vary without a malfunction having occurred.
Incremental vacuum
sampling
Incremental vacuum sampling helps you troubleshoot the problem. The
individual steps of vacuum sampling can be started one after another and
therefore each individual step can be checked.
A read-only menu apart from:
Distributor position
Select which bottle should be filled with the sample.
11.10 Resetting the device
‣
Select a device restart or the factory default setting under
Menu
/
System
/
Device
restart
or
Factory default
.
Parameter description
Function
Info
Device restart
Restart and keep all the settings
Factory default
Restart with factory settings. Settings that have not been saved are lost.
11.10.1 Operating time information
The following information is displayed:
•
Operating hours device
:
Displays the total operating hours of the device in days, hours and minutes
•
Operating hours cooling
:
Displays the total operating hours of the compressor in days, hours and minutes
•
Overfill sensor
(for version with vacuum pump):
Number of times a safety switchoff has been caused by LF2
•
Dosing valve
(for version with vacuum pump):
Number of times the dosing valve is actuated;
‐
> corresponds to the number of samples
taken
•
Vacuum pump
(for version with vacuum pump):
Displays the pump operating time in hours and minutes
•
Sample totalizer
(for version with peristaltic pump):
Number of all samples taken and sample errors
•
Pump tube life
(for version with peristaltic pump):
Displays how old the tube is in days, hours and minutes
•
Peristaltic pump
(for version with peristaltic pump):
Displays the pump operating time in hours and minutes
This counter must be reset when a tube is replaced.
With
Reset
, you reset the counter to zero.
11.10.2 Status of inputs/outputs
Path:
MEAS
/
Measurement