Diagnostics and troubleshooting
Hauser
49
Parameter description
Function
Info
Show
Select a particular event to display more detailed information.
Go to date
Use this function to go directly to a specific time in the list. In this way, you avoid
having to scroll through all the information. The complete list is always visible,
however.
Show summary of inputs
The configured counters of the current or binary input are displayed. Max. 8 lines
11.8.5 Bottle filling
Display shows the bottle filling of the sampler:
‣
In
Menu
/
Diagnostics
select the
Show summary of current program
item or select the
MEAS soft key; works both when the program is active and when it has stopped.
an overview of bottle filling appears for each individual bottle when the program is
started. This gives you detailed feedback on the last sampling operations.
The bottle filling is deleted when the following event occurs:
Program start
The bottle filling is selectively overwritten when the following event occurs:
When the first bottle is reached in situations where "Continuous operation" is
configured as the end of the program in the program settings
The bottle filling is displayed as follows:
A0045690
Display
Info
bt
The bottle number is displayed.
hh:mm
The time the first sample was transferred to the bottle is displayed.
DD-hh:mm
The time the first sample was transferred to the bottle is displayed.
Smp
Displays how often sampling was triggered per bottle.
n.s
Indicates the number of times a sample was not taken even though sampling was
triggered. This can occur if the maximum permissible fill volume for the bottle has
been reached but the system is still supposed to transfer samples to the bottle. The
"Overfill sensor" message is displayed while the program is active.
n.f
The value indicates how often sampling was canceled because it was not possible to
draw in any or enough medium into the dosing glass to cover the LF1 probe.
ml
The sampling volume collected per bottle is displayed.
Q
The total flow for every bottle is displayed (only if flow measurement is connected).