Liquistation CSF28
Electrical connection
Hauser
23
1)
Internal device connection, do not disconnect the plug.
6.4
Connecting the supply voltage
6.4.1
Laying the cable
‣
Lay the cables so that they are protected behind the rear panel of the device.
• Cable glands (up to 8 depending on the version) are available for the cable entry.
• The cable length from the foundation to the terminal connection is approx.
1.7 m (5.6 ft).
• For analyzer stands, the cable length is approx. 1.8 m (5.9 ft) from the foundation.
6.4.2
Cable types
• Power supply: e.g. NYY-J; 3-wire; min. 2.5 mm
2
• Analog, signal and transmission cables: e.g. LiYY 10 x 0.34 mm
2
The terminal connection is protected under an additional cover in the top rear section of
the device.
‣
Therefore remove the rear panel of the device to connect the power supply before
commissioning.
6.4.3
Removing the rear panel of the dosing compartment
1. Open the dosing compartment door.
2. Using an 5 mm (0.17 in) Allen key, release the rear panel by turning the lock
clockwise.
CLOSE
OPEN
1
A0012803