Liquistation CSF28
Electrical connection
Hauser
17
6
Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
NOTICE
The device does not have a power switch
‣
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the
local regulations for installation.
‣
The circuit breaker must be a switch or power switch, and you must label it as the
circuit breaker for the device.
‣
The protective ground connection must be established before all other connections. If
the protective ground is disconnected, this can be a source of danger.
‣
A circuit breaker must be located near the device.
NOTICE
The device does not have a power switch
‣
The mains switch can be ordered via a TSP modification.
‣
A fuse with a maximum rating of 10 A must be provided internally when operating
with a power cable. The fuse can be mounted under the rear cover.
‣
The protective ground connection must be established before all other connections. If
the protective ground is disconnected, this can be a source of danger.