Mounting
Liquistation CSF28
16
Hauser
Connecting the suction line on version with peristaltic pump
1
2
3
4
A0013705
9
Connecting the suction line from the side (as-delivered
state)
1
Small thread adapter nut
2
Hose
3
Thread adapter nut
4
Hose gland
1
A0013706
10
Suction line connected from below
Modification of the suction line from side connection to connection from below
1. Unscrew the thread adapter nut (item 3) and the hose gland (item 4) from the side
panel.
2. Unscrew the small thread adapter nut (item 1) and remove the hose.
3. Fit a hose adapter on the hose.
4. Secure the hose adapter with a worm drive hose clip.
5. Connect the suction line from below as illustrated.
6. Insert the dummy plugs supplied.
5.3
Post-installation check
1. Verify that the suction line is securely connected to the device.
2. Visually check that the suction line is installed correctly from the sampling point to
the device.
3. Verify that the distribution arm is correctly engaged.
4. Allow the sampler to rest for at least 12 hours after setup and before switching it on.
Otherwise you may cause damage to the climate control module.