Mounting
Liquiphant FTL62 Density
10
Hauser
A0042281
2
Handling the device during transportation
5
Mounting
L
WARNING
Loss of protection rating if the device is opened in a wet environment.
‣
Only open the device in a dry environment!
Mounting instructions
• Any orientation for device with short pipe up to approx. 500 mm (19.7 in)
• Vertical orientation from above for device with long pipe
• Minimum distance between the fork tip and the tank wall or pipe wall: 10 mm (0.39 in)
1
A0048473
3
Installation examples for a vessel, tank or pipe
1
Vessel insulation (example with temperature spacer/pressure-tight feedthrough)
If process temperatures are high, the device should be included in a vessel insulation system to prevent the
electronics from heating as a result of thermal radiation or convection.
5.1
Mounting requirements
For density measurement, the tuning fork must always be completely immersed.