24
Cond
Measurement
CAL
CELL CONST.
The conductivity will be shown in the display
and can be compared with a reference solution
(temperature‑corrected).
Value blinks
Use
st
to select the value for the cell constant.
Calibration will be performed. Automatic return to measuring mode.
CELL CONST. Calibration
(Calibration by entry of cell constant)
The calibration method is selected in the configuration menu.
Measurement
CAL
COND
Dip sensor in solution.
Value blinks
Use
st
to adjust the temperature‑corrected
conductivity value. NOTICE: Here, the meter
does not perform a temperature compensation!
Calibration will be performed. Automatic return to measuring mode.
COND Calibration
(Calibration by entry of conductivity)
The calibration method is selected in the configuration menu.
Conductivity Calibration