32
pH
Oxy Cond
Measuring
Once you have completed all preparations, you can start with the actual measure-
ment.
1) Connect the desired sensor to the meter. Some sensors require a special prepara-
tion. Please proceed according to the operating instructions for the sensor.
2) Switch the device on using the
or
key.
3) Depending on the measurement method and the sensor used, immerse the
sensing part of the sensor in the medium to be measured.
4) Watch the display and wait for the reading to stabilize.
Toggling the Measured Value Display
During measurement, you can toggle between display of primary / secondary
measured value and clock by pressing
.
Adjusting the Temperature
When you connect a sensor without temperature detector, you can manually adjust
the temperature for measurement or calibration:
1) Press
to access measuring mode. The adjusted temperature will be displayed.
2) Set the desired temperature value using the
or
arrow.
Holding the key depressed changes the temperature value at high speed.