Oxy
Oxygen Calibration
1) In measuring mode, press the Menu softkey.
2) Select “Calibration” and confirm by pressing E.
3) Select desired “Calibration mode” and confirm by pressing E.
4) Select “Membrane module replacement” if you wish to save a change of
membrane or electrolyte in the connected sensor. The digital, optical oxygen
sensor automatically recognizes when its sensor cap has been replaced.
5) Select “TAG” to enter a measuring point which is to be saved in the sensor
together with the calibration data record.
6) Perform the selected calibration as described on the following pages.
Follow the instructions on the display.
Calibration in Air
(Calibrating the slope in air)
1) Place sensor in air and wait for a stable measured value.
2) Press Start softkey.
3) Adjust the correct value for “Relative humidity“. Then press Continue.
Calibration will be performed.
4) Finally, you can Apply or Discard these values.
Note: To abort calibration, you can press
at any time.