28
Cond
Calibration
Conductivity Calibration
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing E.
3) Select the desired “Calibration mode”.
4) Perform the selected calibration as described on the following pages.
Follow the instructions on the display.
Auto Calibration
(Automatic calibration with specification of the calibration solution used)
5) Select the calibration solution:
5 °C
10°C
15 °C
20 °C
25 °C
30 °C
35 °C
40 °C
50 °C
CLY 11-A
46.4
52.9
59.7
66.7
74.0
81.4
88.8
96.5
112.2 µS/
cm
CLY 11-B
93.9
107.1
120.8
135.2
149.6
164.5
179.7
195.1
226.8
CLY 11-C
0.8
1.010
1.136
1.270
1.406
1.542
1.683
1.824
2.114
mS/
cm
CLY 11-D 8.07
9.16
10.29
11.45
12.64
13.86
15.11
16.37
18.97
CLY 11-E 70.58 79.34
88.20
97.56 107.00 116.52 126.10 135.98 155.82
6) Press the Start softkey.
7) Immerse the sensor in the solution and press Continue.
8) Finally, the calibration data record will be displayed.
You can Apply or Discard these values.
NOTICE!
• Make sure that the values of the calibration solutions used corre-
spond exactly to those specified in this manual. If not, the resulting
cell constant will be incorrect.
• When calibrating in a liquid, make sure that the sensor, the separate
temperature probe (if present) and the calibration solution have
the same temperature. Only this ensures that the cell constant is
determined correctly.
Note: To abort calibration, you can press
at any time.