Operation
Liquiline System CA80HA
78
Hauser
1. Insert the SD card into the controller card reader.
2. Go to
Menu
/
Setup
/
General settings
/
Extended setup
/
Data management
/
Export
setup
.
3.
Name
: Assign a file name.
4. Then select
Export
.
If you have already assigned the file name, you will be asked whether you want to
overwrite the existing setup.
5. Use
OK
to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the
Save setup
.
This is the only way you can save a setup to an SD card and reload it later on or upload it
to other devices.
Activation code
You require activation codes for:
• Additional functions, e.g. fieldbus communication
• Firmware upgrades
If your original device has activation codes, these can be found on the nameplate. The
corresponding device functions are activated at the factory. You only require the codes if
servicing the device.
1. Enter the activation code:
Menu
/
Setup
/
General settings
/
Extended setup
/
Data
management
/
Activation code
.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function
Activation code beginning with
Two current outputs (BASE2-E module only)
081...
Web server
351...
PROFIBUS DP
0B3...
Modbus TCP
0B8...
Modbus RS485
0B5...
EtherNet/IP
0B9...
PROFINET
0B7...
Cooling
0F1...
Feedforward control
220...
Chemoclean Plus
25...
Collecting vessel
20...