Operation
Liquiline System CA80HA
60
Hauser
• Device tag
Give your device any name of your choice (max. 32 characters).
• Set date
Correct the set date if necessary.
• Set time
Correct the set time if necessary.
2. Insert the bottles and activate the bottles used in the menu:
Bottle insertion
/
Bottle
selection
.
3. Check the concentration of the calibration standard used:
Calibration
/
Settings
/
Nominal concentration
.
4. Optionally, also change the measuring interval:
Measurement
/
Measuring interval
.
All the other settings can be left in the default factory settings for the time being.
5. Return to the measuring mode: press and hold the soft key for
ESC
for at least one
second.
Your analyzer now works with your general settings. Optionally connected sensors
use the factory settings of the specific sensor type and the individual calibration
settings that were last saved.
If you want to already configure additional input and output parameters in the
Basic setup
analyzer
:
‣
Configure the current outputs, relays, limit switches and device diagnostics with the
following submenus.
10
Operation
10.1 Display
10.1.1 Soft keys in the measuring mode
On the bottom row of the display you can find four soft keys in the measuring screens:
• With
MENU
and
DIAG
take you directly to the specific software menu.
• With
HOLD
you can activate an immediate, general hold for sensors. This also sets all the
linked outputs, controllers and cleaning cycles to HOLD. Any sensor cleaning programs that
are currently running are interrupted. However, you can also start manual sensor cleaning
when a hold is active.
• With
MODE
takes you to a picklist of frequently used software functions.