Liquiline System CA80HA
Commissioning
Hauser
59
L
WARNING
Connection errors
The safety of people and of the measuring point is under threat. The manufacturer does not
accept any responsibility for errors that result from failure to comply with the instructions in
this manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Device condition and specifications
‣
Are the hoses free from damage on the outside?
Visual inspection of the liquid-bearing lines
‣
Check the hose connections using the hose connection diagram.
‣
Is the suction line connected to the sample collector (if present)?
‣
Are the dispensers inserted correctly?
‣
Can the dispensers move up and down freely?
‣
Are all the hose connections leak-tight?
‣
If sample preparation is provided: has the connection been made? Are the protective hoses
in the hose glands strain-relieved?
‣
If sample preparation is not connected: is the sample hose in the hose gland strain-
relieved?
‣
Have the bottles with reagents, and standard been inserted and connected?
9.3
Switching on the measuring device
1. Connect the power supply.
2. Wait for the initialization to finish.
9.4
Setting the operating language
Configuring the language
1. Press the soft key:
MENU
.
2. Set your language in the top menu item.
The device can now be operated in your chosen language.
9.5
Configuring the measuring device
9.5.1
Basic setup analyzer
Making basic settings
1. Switch to the menu
Setup
/
Basic setup analyzer
.
Make the following settings.