System integration
Liquiline System CA80HA
46
Hauser
3. Narrow your search by specifying the driver type, product root and communication
protocol.
4. Search.
5. Click the suitable entry from the list of hits.
You receive the download links including an extended description.
7.2
Integrating the measuring device into the system
7.2.1
Web server
Versions without a fieldbus: you require an activation code for the web server.
Connection
‣
Connect the communication cable of the computer to the Ethernet port of the BASE2
module.
BASE2
Web Browser
A0044872
51
Web server/Ethernet connection
Establishing the data connection
All versions with the exception of PROFINET:
You must switch off the
DHCP
parameter in the Ethernet settings for your device to have a
valid IP address. (
Menu
/
Setup
/
General settings
/
Extended setup
/
Ethernet
/
Settings
)
You can assign the IP address manually in the same menu (for point-to-point connections).
All versions including PROFINET:
The IP address and subnet mask of the device can be found under:
DIAG
/
System
information
/
Ethernet
.
1. Start your PC.
2. First, configure a manual IP address in the network connection settings of the operating
system.