Liquiline System CA80HA
Diagnostics and troubleshooting
Hauser
173
DIAG/Simulation
Function
Options
Info
Meas. inputs
Simulation of a measured value (only for
sensors)
This menu appears once for each measuring
input.
Channel : parameter
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the measured value, this is
indicated on the display by a simulation icon in
front of the measured value.
Main value
Depends on the sensor
‣
Set the desired simulation value.
Sim. temperature
Selection
• Off
• On
Factory setting
Off
If you simulate the temperature measured
value, this is indicated on the display by a
simulation icon in front of the temperature.
Temperature
-50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
‣
Set the desired simulation value.
Binary input x:y
Binary output x:y
Simulation of a binary input or output signal
The number of submenus available
corresponds to the number of binary inputs or
outputs.
Simulation
Selection
• Off
• On
Factory setting
Off
State
Selection
• Low
• High
11.13 Device test
11.13.1 Analyzer
DIAG/System testAnalyzer/Photometer
Function
Options
Info
Cleaning factor
Read only
Raw value
Read only
Temperature
Read only