Diagnostics and troubleshooting
Liquiline System CA80HA
172
Hauser
11.12 Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
• Relay contact opening or closing
Only current values are simulated. Via the simulation function, it is not possible to
calculate the totalized value for the flow or rainfall.
‣
Before simulation: Enable the inputs and outputs in the Setup menu.
DIAG/Simulation
Function
Options
Info
Current output x:y
Simulation of an output current
This menu appears once for each current
output.
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the value at the current output,
this is indicated on the display by a simulation
icon in front of the current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
‣
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the relay state, this is indicated
on the display by a simulation icon in front of
the relay display.
State
Selection
• Low
• High
Factory setting
Low
‣
Set the desired simulation value.
The relay switches in accordance with your
setting when you switch on the simulation. On
the measured value display, you see
On
(=
Low
) or
Off
(=
High
) for the simulated relay
state.