Operation
Liquiline System CA80CR
98
Hauser
10.6 Outputs
10.6.1 Current outputs
Liquiline System CA80 has two analog current outputs as standard.
Set the current range from 0 to 20 mA or 4 to 20 mA under Menu/Setup/General settings.
Possible applications
• For outputting a measured value to a process control system or an external recorder
• For outputting a manipulated variable to connected actuators
The current output curve is always linear.
Operating mode
Function = "Diagnostics mes-
sage"
Options
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
With this selection, the diagnosis messages are trans-
mitted over the binary output that you individually
allocated to it.
Namur M ... F
When you choose one of the Namur classes, all messa-
ges allocated to that respective class are output.
You can also change the Namur class assignment for
every diagnostics message.
(Menu/Setup/General settings/Diagnostics/Device
behavior or Menu/Setup/Inputs/../Diagnostics set-
tings/Diag. behavior)
1)
x:y = slot no. : input number
Path: Menu/Setup/Outputs/Binary output x:y
1)
Function
Options
Info