Liquiline System CA80CR
Operation
Hauser
85
10.2.4 Cleaning
Date
01.01.1970 ... 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24 h)
Calibration interval
00-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval of the calibra-
tion/adjustment.
Next calibration
Mode = "Automatic"
Read only
Zero point
Read only
Calibration factor
Read only
Relationship of the measured concentration to the
pre-specified concentration of the calibration standard.
Settings
Nominal concentration
0.5 to 4.5 (mg/l)
Factory setting
2.0 (based on Cr(VI))
For configuring the concentration of the calibration
standard solution.
Depends on the settings under Menu/Setup/Analy-
zer/Extended setup/Measuring value/Main value
Path: Menu/Setup/Analyzer/Cleaning
Function
Options
Info
Start condition
Options
• Immediate
• Date/time
• Disabled
Factory setting
Disabled
Immediate:
Cleaning of starts immediately.
Date/time:
Cleaning of starts at the selected date/time.
Disabled:
cleaning is not enabled.
If Start condition
Date/time
is selected:
Date
01.01.1970 ... 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24 h)
Path: Menu/Setup/Analyzer/Calibration
Function
Options
Info