Liquiline System CA80CR
Maintenance
Hauser
153
12.4 Changing the hose
You need the following parts:
1. Rinse the system (see the "Rinsing the system" section).
There should not be any sample left in the sample collecting vessel.
2. Remove the bottles and the bottle tray.
3. Replace with hoses of the same diameter and length. Attach hose labels to the new hoses.
The drain hose D6 must be routed behind syringe 7.
4. place the bottle tray back into the housing.
5. Connect the bottles to the appropriate covers and hoses.
6. Select "Menu/Operation/Maintenance/Bottle replacement/Bottle insertion/Bottle
selection".
7. Select all the bottles and press "OK" to confirm.
8. Select "Confirm bottles inserted".
9. Press the soft key "MODE" and select "Continue automatic mode" or "Start automatic mode".
12.5 Changing the filter mats
You need the following parts:
• Filter mats (part of the CAV800 maintenance kit)
1. Open and remove the fan screen on the right and left underside of the analyzer.
Recommendation for the hose change interval:
• NORPRENE, black
Every 12 months
• C-Flex, white, ID 1.6 mm Every 12 months
• C-Flex, white, ID 3.2 mm if necessary (every 2 to 3 years)
• C-Flex, white, ID 6.4 mm if necessary (every 2 to 3 years)
• NORPRENE hose, ID 1.6 mm
Part of the CAV800 maintenance kit
• C-Flex hose, ID 1.6 mm
• C-Flex hose, ID 3.2 mm
• C-Flex hose, ID 6.4 mm
• Hose fitting
• 1 pair of gloves that are resistant to the reagents used