Diagnostics and troubleshooting
Liquiline System CA80AM
150
Hauser
11.6.2 Sensor-specific diagnostics messages
See Operating Instructions for "Memosens", BA01245C
11.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status.
Furthermore, various service functions are available.
The following messages are directly displayed every time you enter the menu:
• "Most important message"
Diagnostics message recorded with the highest criticality level
• "Past message"
Diagnostic message whose cause is no longer present.
All the other functions in the Diagnostics menu are described in the following chapters.
11.8 Diagnostics list
All the current diagnostics messages are listed here.
A time stamp is available for each message. Furthermore, the system also displays the
configuration and description of the message as saved in "Menu/Setup/General
settings/Diagnostics/Device behavior".
971 Current Input low
S
On
On
Current input too low
At 4 to 20 mA, the input current is less than the
lower error current.
--> Check the input for short-circuits.
972 Curr. input > 20 mA
S
On
On
Current output range exceeded
973 Current Input < 4 mA
S
On
On
Current output range undershot
974 Diagnostics confirm.
C
Off
Off
User has acknowledged the message displayed in
the measuring menu.
975 Device reset
C
Off
Off
Device reset
978 ChemoClean Failsafe
S
On
On
No feedback signal detected within the
configured period.
1. Check the application
2. Check the wiring
3. Extend the duration
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current