Liquiline System CA80AL
Commissioning
Hauser
49
9
Commissioning
9.1
Preparatory steps
Prior to connecting the device to the power supply
On account of the device design, high switch-on currents occur when the device is
commissioned at low temperatures. The power value indicated on the nameplate refers to
the power consumption after one minute of operation when the device is commissioned at
5 °C (41 °F) .
1. To avoid overloading the power cable or triggering the mains fuse, only commission
the device at temperatures ≥5˚C (41 °F).
2. Connect the liquid-bearing pipes and visually inspect the hoses.
L
CAUTION
Automatic cleaning function for the suction hose
Risk of injury from contact with very acidic cleaning solution
‣
Do not shorten the system intake hose.
L
CAUTION
Activities while the analyzer is in operation
Risk of injury and infection from medium, reagents or cleaner!
‣
Before you release any hoses, make sure that no actions, such as the pumping of
sample, are currently running or are due to start shortly.
‣
Wear protective clothing, goggles and gloves or take other suitable measures to protect
yourself.
‣
Wipe up any spilt reagent with a disposable tissue and rinse with clear water. Then dry
the cleaned areas with a cloth.
9.1.1
Hose connection diagram
A0041298
66
Hose connection diagram
The diagrams below reflect the status at the
time of issue of this documentation. The
hose connection diagram that applies for
your device version is provided on the inside
of the door of the analyzer.
‣
Only connect the hoses as specified in
this diagram.