Liquiline CM44P
Electrical connection
Hauser
43
Example: Connecting the cleaning unit 71072583 for CAS40D
NOTICE
Power consumption too high for the Liquiline alarm relay!
Can cause irreparable damage to the base module
‣
Connect the cleaning unit only to terminals of an additional module (AOR, 2R or 4R),
not
to the alarm relay of the base module.
PE
L1
N
PE
T2.0A
230 V AC / 50 Hz
1
41
43
42
Liquiline/Sampler/
Analyzer:
AOR/2R/4R
A0028597
61
Connecting the cleaning unit for CAS40D
Example: Connecting the Chemoclean CYR10 injector cleaning unit
41
43
42
Relay 1
41
43
42
Relay 2
1
2
1
2
N
L
PE
CM44x/CM44xR/
CM44P
CYR10
1
2
3
4
5
A0028598
62
Connecting the CYR10 injector cleaning unit
1
External power supply
2
Cleaner to spray head
3
Container with cleaner
4
Motive water 2 to 12 bar (30 to 180 psi)
5
Backflow valve (to be provided by the customer)