Accessories
Levelflex FMP56, FMP57 FOUNDATION Fieldbus
122
Hauser
16.1.5 Remote display FHX50
Accessory
Description
Remote display FHX50
A0019128
• Material:
– Plastics PBT
– CF3M (similar to 316L/1.4404)
– Aluminum (in preparation)
• Ingress protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for the display modules:
– SD02 (push buttons)
– SD03 (touch control)
• Connection cable:
– Cable with M12 plug; supplied with the FHX50; up to 30 m (98 ft)
– Customer supplied standard cable; up to 60 m (196 ft)
• Ambient temperature: –40 to 80 °C (–40 to 176 °F)
• If the remote display is to be used, the device must be ordered in the version "Prepared for display FHX50"
(feature 030, option L or M). For the FHX50, on the other hand, the option A: "Prepared for display FHX50" has
to be selected in feature 050: "Option Measurement Device".
• If a device has not been ordered in the version "Prepared for display FHX50", but is nevertheless to be equipped
with an FHX50, it is essential to select the option B: "Not prepared for display FHX50" in feature 050: "Option
Measurement Device" of the FHX50. In this case, a retrofit kit, needed to prepare the device for the remote
display, is supplied together with the FHX50.
For transmitters with approval, application of the FHX50 may be restricted. A device may only be retrofitted with
the FHX50 if option L or M ("Prepared for FHX50") is quoted under
Basic specifications
, position 4 "Display,
operation" in the associated Safety Instructions (XA).
In addition to this, observe the Safety Instructions (XA) of the FHX50.
Do not retrofit transmitters with:
• approval for use in areas with combustible dusts (Dust-Ex approval)
• type of protection Ex nA
For details refer to the document SD01007F.